已注册用户需要常规用户名密码登录并绑定微信,才可以扫码
点击右上角可切换登录模式
还没有帐号?
现在注册并享受以下权益:
- 直达官方商城在线购买
- 请求报价
- 参与精彩问答
- 获取会员积分
- 兑换礼品
- 和更多 ...
SP510 Maintenance Requirements
SP510,水质硬度检测仪
Colorimeter needs to be cleaned monthly or as needed
Reagents
Replace buffer and indicator reagents every two months.
Discard old containers and any unused contents.
Standardization
Standardization must be performed after reagent installation or when an optical system component has been replaced.
Replace pump tubing
Every 6 months < 27oC
Every 3 months > 27oC
Replace remaining instrument tubing annually
隐私政策 Cookie通知 / 沪ICP备13034148号-4 / 沪公网安备31010502004971号 / 沪(浦)应急管危经许[2023]201871
© 版权 2015-2024 哈希中国版权所有 / 哈希水质分析仪器(上海)有限公司 / 本网站用字经北大方正电子有限公司授权许可