SP510 Maintenance Requirements
SP510,水质硬度检测仪
Colorimeter needs to be cleaned monthly or as needed
Reagents
Replace buffer and indicator reagents every two months.
Discard old containers and any unused contents.
Standardization
Standardization must be performed after reagent installation or when an optical system component has been replaced.
Replace pump tubing
Every 6 months < 27oC
Every 3 months > 27oC
Replace remaining instrument tubing annually